Performance data in serum

Refer to package insert for detailed information on serum sample preparation.

Linearity

Figure 1: Linearity between 1.7 µmol/L up to 200 µmol/L total bile acids (red dashed line represents the 10% limits)

Method comparison

Figure 2: Method comparison DiaSys vs. competitor

Precision

Table 1: Repeatability of DiaSys Total bile acids 21 FS

Mean [µmol/L] SD [µmol/L] CV [%]
Sample 15.410.1292.38
Sample 210.20.0850.829
Sample 31991.190.598

Table 2: Reproducibility of DiaSys Total bile acids 21 FS

Mean [µmol/L]SD [µmol/L]CV [%]
Sample 15.390.0791.42
Sample 210.40.1591.54
Sample 3 2011.640.816