Performance data in serum
Refer to package insert for detailed information on serum sample preparation.
Figure 1: Linearity between 1.7 µmol/L up to 200 µmol/L total bile acids (red dashed line represents the 10% limits)
Figure 2: Method comparison DiaSys vs. competitor
Precision
Table 1: Repeatability of DiaSys Total bile acids 21 FS
| Mean [µmol/L] | SD [µmol/L] | CV [%] | |
|---|---|---|---|
| Sample 1 | 5.41 | 0.129 | 2.38 |
| Sample 2 | 10.2 | 0.085 | 0.829 |
| Sample 3 | 199 | 1.19 | 0.598 |
Table 2: Reproducibility of DiaSys Total bile acids 21 FS
| Mean [µmol/L] | SD [µmol/L] | CV [%] | |
|---|---|---|---|
| Sample 1 | 5.39 | 0.079 | 1.42 |
| Sample 2 | 10.4 | 0.159 | 1.54 |
| Sample 3 | 201 | 1.64 | 0.816 |